IWA-FP08 Wired X-Ray Flat-Panel Detector (Venu 1717XV, 17×17 inch)

IWA-FP08 Venu 1717XV is a wired 17×17 inch static flat-panel detector manufactured by IRAY (Shanghai iRay Technology). The detector is built around an amorphous-silicon (a-Si) TFT array with a CsI scintillator and delivers a 2912 × 2912 pixel matrix at 139 µm pitch with 3.6 LP/mm spatial resolution.

This detector is suited for: Fixed-room veterinary radiography with stable Ethernet connectivity, retrofitting wall-stand and table bucky systems without wireless network requirements.

Key Features

  • 17×17 inch active imaging area (43 × 43 cm) with 2912 × 2912 pixel matrix at 139 µm pitch.
  • 3.6 LP/mm spatial resolution with a-Si TFT photodiode + CsI scintillator for high DQE.
  • 16-bit A/D conversion for full dynamic range.
  • Trigger modes: AED / Prep / Software — compatible with any X-ray generator.
  • Gigabit Ethernet data interface.
  • Wired-only configuration — no battery management or charging cycle to maintain.
  • Preview image: 3 s, cycle time: 5 s.

Technical Specifications

Brand IRAY (Shanghai iRay Technology)
Model Venu 1717XV
Detector technology Amorphous silicon (a-Si) TFT
Scintillator CsI
Active imaging area 43 × 43 cm (17×17 inch)
Pixel matrix 2912 × 2912
Pixel pitch 139 µm
Spatial resolution 3.6 LP/mm
A/D conversion 16 bit
Trigger modes AED / Prep / Software
Preview image time 3 s
Cycle time 5 s
Wireless / wired Wired (Ethernet)
Data interface Ethernet (GigE)
Dimensions 460 × 460 × 15 mm
Weight 4.0 kg
Operating temperature 5 – 35 °C
Operating humidity 10 – 90 % RH (non-condensing)

Clinical Applications

  • Veterinary general radiography (thoracic, abdominal, skeletal)
  • Fixed-room veterinary radiography with stable Ethernet connectivity, retrofitting wall-stand and table bucky systems without wireless network requirements.
  • Replacing legacy CR cassettes in mixed-practice clinics
  • Fixed-room DR workflow with stable Ethernet connectivity

Workflow & Installation

The detector connects to a host PC through Gigabit Ethernet. IRAY supplies the iDetector / iRayDR acquisition software for first-run setup, calibration and routine maintenance. Calibration is recommended every 6 months or when X-ray generator / tube / collimator components are replaced. Cleaning: wipe with a non-abrasive surface disinfectant; do not immerse.

Package Contents

  • 1 × IRAY Venu 1717XV flat-panel detector
  • 1 × Detector cable (Ethernet)
  • 1 × Power adapter
  • 1 × iDetector / iRayDR acquisition software
  • 1 × User manual
  • 1 × Quality certificate

Downloads

Why Choose IWA-FP08

  • Workflow efficiency — wired Ethernet provides the most stable connectivity for fixed-room workflows.
  • Clinical accuracy — 3.6 LP/mm spatial resolution and 16-bit A/D with CsI scintillator deliver diagnostic-grade image quality.
  • Manufacturer reliability — IRAY is a leading global X-ray detector OEM with extensive worldwide installation base.

Warranty & Support

Manufacturer warranty per supply agreement. Spare parts and authorised in-region service available through IWA Medical / LumaVet.

FAQ

Q: Is IWA-FP08 a static or dynamic detector?
Static. The Venu 1717XV is designed for single-frame radiographic exposures (not fluoroscopy or dynamic imaging).

Q: What X-ray generators is it compatible with?
Any conventional X-ray generator. The detector supports AED (Auto Exposure Detection) for use without hard-wired sync, plus software-trigger workflow.

Q: Does it support DICOM?
The IRAY iDetector / iRayDR acquisition software outputs raw and processed images; DICOM 3.0 integration is typically provided by the host DR workstation software bundled with iVision / iVet / iPet workstations or by your existing PACS gateway.